| 1. Match the optical method | Confirm the sensor’s light source and the method required by your monitoring program. | ISO 7027-1 nephelometric measurement uses near-infrared light at 860 nm. U.S. EPA Method 180.1 uses a tungsten lamp with a specified color temperature of 2,200–3,000 K and a detector response in the visible-light region. | Choose a sensor designed for the applicable method. Readings from different optical methods may not agree for every sample, especially when particle size or color varies. |
| 2. Check the measurement geometry | Compare the detector angle and measurement configuration with the method and application. | Both methods use nephelometric detection at approximately 90° to the incident light beam. ISO 7027-1 specifies a detector angle of 90° ± 2.5° for its nephelometric method. | For method-based reporting, verify the instrument documentation rather than assuming all sensors with a 90° detector are method-equivalent. |
| 3. Select an appropriate measuring range | Consider expected minimum, normal, and peak turbidity, as well as the required resolution. | Usable range and resolution depend on the instrument design, optical path, sample, and measurement method; they are not determined by wavelength alone. | Choose a range that covers expected peaks without sacrificing useful resolution at routine levels. Check the specified range and performance under the conditions in which the sensor will operate. |
| 4. Confirm the reporting units | Check whether the application requires NTU, FNU, or another method-defined reporting convention. | NTU is commonly associated with EPA Method 180.1, while FNU is commonly used for ISO 7027 infrared nephelometric measurements. | Do not treat NTU and FNU as automatically interchangeable. Record the method and units, and establish any comparison using the actual water matrix and instruments involved. |
| 5. Evaluate sample color and particle characteristics | Consider dissolved color, suspended-particle size, shape, and composition. | Optical response can vary with the sample. Visible-light and 860 nm systems may respond differently to colored water and to particles with different scattering properties. | If samples are colored or vary substantially in particle composition, test candidate sensors on representative samples and compare results with the required reference method. |
| 6. Review calibration and verification needs | Check the calibration procedure, standards, frequency, and records required for the application. | Formazin-based standards are used in established turbidity measurement procedures, including EPA Method 180.1. Follow the applicable method and the sensor’s instructions for calibration and verification. | Use suitable, traceable standards and document calibration, checks, cleaning, and any corrective actions. Do not assume that a calibration for one optical method validates another. |
| 7. Match the sensor to the installation | Assess mounting, sample flow, bubbles, fouling, cleaning access, and maintenance requirements. | Bubbles, deposits on optical surfaces, and changing flow or sample conditions can affect optical measurements, regardless of the selected wavelength. | Choose a configuration that can be installed and maintained reliably at the measurement point. Plan routine inspection and cleaning, and verify performance in the installed setup. |